ISSN: 2641-1709
Petrov SM*
Received: December 14, 2021; Published: January 06, 2022
Corresponding author: Petrov SM, PE, Saint Petersburg, Russia
DOI: 10.32474/SJO.2022.07.000275
Frequency range (FR) of the cochlear implantation (CI) system is one of the main parameters of the CI speech processor. FR is set at the first connection and may change (according to indications) during the use of the CI. In implants of different companies, it is possible to set FR in different boundaries - from 70 Hz to 10 kHz, but since the main goal of CI is successful speech communication, the issue of setting the boundaries of FR is considered in this paper specifically in relation to the perception of a speech signal. What frequency range is the best range for speech perception by a subject with CI? In this article, we rely on the parameters of the Med-El implants, but all the above considerations are applicable to the CI systems of any companies. The issue of disconnection of failed and short-circuited electrodes are discussed.
Keywords: Cochlear implantation; speech intelligibility; frequency range; channel selectivity of stimulation
The frequency range (FR) of the cochlear implantation (CI)
system, i.e., the frequency range processed by the implant is one
of the main parameters of the CI speech processor. To solve the
issue of setting optimal FR boundaries, it is necessary to consider
how the speech (acoustic) signal is processed by CI, what and how
the CI patient hears as a result of processing and to correlate the
perceived signal with the spectral characteristics of speech. Let’s
briefly consider how the CI processes an acoustic signal. At each
moment of time, the speech signal has a certain instantaneous
(momentary) spectrum. This instantaneous spectrum within
the boundaries of the FR of the implant is divided in the speech
processor into a limited number of frequency bands of different
(defined) widths in accordance with the number of electrodes
(in different CI models from 12 to 24). After processing, they are
converted into electrical pulses. Each of these impulses stimulates
a certain part of the auditory nerve corresponding to the position of
the electrode in the cochlea and each of them causes some colored
auditory sensation, in different channels – from different electrodes
- differing in spectral sensation. The width of the bands into which
the speech spectrum is divided determines the channel selectivity
of stimulation (CSS) (the wider the FR, the smaller the CSS), and
regardless of the bandwidth processed in this channel, the patient
hears in this channel a signal of the same specific spectrum
(tonotopy of cochlea). The spectrum of the audible sound band
in each channel depends on the flow of current in the volumetric
conductor between the electrode in the cochlea and the ground one
under the skin behind the ear.
The instantaneous spectrum of speech is transmitted by a
series of stimuli supplied from the first to the last electrode (not
all electrodes can work in one series, but selectively - the n-of-m
strategy). Thus, the implanted patient perceives the instantaneous
spectrum of speech as sensations from a certain number of
frequency bands. The patient perceives the picture of speech in
time as a change in sensations from the change of pictures of the
instantaneous spectrum. It is obvious that the normal speech
spectrum changes significantly in implanted patients, but what
is important: There are speech signs in the new - transformed
- spectrum. The spectral redundancy of the speech signal
(redundancy means that it can be removed from it without prejudice
to its understanding) is so great [1] that the speech signs that are
present in the speech signal after processing by the processor are
quite enough for the implanted patient to fully master speech.
The main task solved by an audiologist when setting up a speech
processor is to maximally convey the information that is available
in the transformed speech signal and is transmitted using electrical
stimuli directly to the auditory nerve. For an implanted patient,
every bit is expensive. That is why it is of great importance to install
the correct FR. The width of single-channel bands depends on the
FR value. As the FR increases, the width of all single-channel bands
increases. There may be such a situation that some frequency band processed in two channels with a wide FR of the implant will be
processed in three channels with a narrow FR, i.e., with a narrow
FR, the CSS will increase. Naturally, the higher the CSS, the more
accurately the spectrum of processed speech is transmitted by the
implant.
The speech of many European languages has quite similar
spectral characteristics and therefore, for the best perception
(understanding) of speech, general recommendations, and rules
for setting the boundary of the frequency range in all CI systems
can be found. FR is directly related to the content of speech
information in it and the distribution of information between the
channels of the implant. All other things being equal, in order to
obtain maximum speech intelligibility in CI systems, it is necessary
to use the best frequency range. Based on many years of experience
in CI, the author had the idea that in order to improve speech
intelligibility, it is necessary to make some changes to the acoustic
signal processing program in the processor. Let’s consider a speech
signal from the point of view of its processing in CI. Speech signals
are limited to the frequency range of 100-10000 Hz [2]. It has been
established, however, that the quality of speech is quite satisfactory
when the spectrum is limited to frequencies of 300-3400 Hz. These
frequencies are accepted by the International Telecommunication
Union as the boundaries of the effective speech spectrum, i.e.,
this mid-frequency range contains a significant part of speech
information. With this frequency band, the intelligibility of phrases
is more than 99% and the satisfactory naturalness of the sound
is preserved. Naturally, after processing this band of the speech
spectrum in the CI, some of the information is lost and what the CI
patient hears will not be enough for a 100 percent understanding
of speech. It is obvious that for CI patients, these losses must be
compensated by expanding the band of perceived frequencies
of this effective speech spectrum. Question: to what limits in the
direction of low and high frequencies? Is the range from 100 to
10000 Hz suitable? It should be understood that CI is not a hearing
aid (HA). This is in the HA, the wider the frequency range the
patient hears, the better.
To resolve the issue of setting the boundaries of the FR, we must
consider the data on the contribution of low and high frequency
regions of the speech spectrum in speech intelligibility. We consider
the processing of the speech spectrum in a wide FR of the Med-El
implant - 70-8500 Hz. In the study of M.A.Sapozhkov was shown
that the frequency band up to 420 Hz contains 5% of speech
information [3]. This band with a wide FR (70-8500 Hz) of CI is
processed in the first-second and partially in the third channels of
the implant. Obviously, the lower the frequency band of the first,
second and third channels on the frequency scale, the less speech
information it contains. It is obvious that the band of the first channel
contains only a small part of speech information (less than 1%) and
with a wide FR, the first electrode works almost in vain. The band
of the second channel contains slightly more information than the
band of the first channel. It follows that from the first and second
channels only a small part of the audio information comes to the
implant. Therefore, it is not necessary to expand the FR towards the
low frequency to 70 Hz. When considering high frequencies (HF)
in the FR it follows that the band from 6500 to 8500 Hz contains
5% of speech information [3], which partially contains information
about only a few high-frequency phonemes. It must be remembered
that the spectral features of these HF phonemes are also present
in the low and midrange frequencies. It can be assumed that by
narrowing the FR from the HF side, we will lose a small part of the
HF information, but we will increase the CSS and thereby sharpen
the perception of the remaining low-middle frequency phonemes.
Thus, by narrowing the FR of 70-8500 Hz, we lose a small
part of the speech information from the low-frequency and highfrequency
parts of the speech spectrum, but we increase the CSS
in the informative mid-frequency part of the spectrum. So, in this
way we can find the golden mean between the loss of information
in the low and high-frequency zones (with a narrowing of the FR
70-8500 Hz) and increase the CSS by reducing the width of all
processed single-channel bands. From the point of view of the
above considerations, it is interesting to consider the results of our
previous study. In that study, we modeled 4 CI with different FR and
compared the intelligibility of the processed speech. It should be
noted that in this work, we developed a methodology for measuring
the intelligibility of words which was successfully applied. It can also
be used when comparing CI programs with different strategies. For
further discussion, we will present a table of the results obtained
in the previous study [4]. The averaged results of measurements
of the intelligibility of spectrally deprived words depending on the
width of the frequency range of the CI are presented in Table 1.
Table 1: Intelligibility of spectrally deprived words (%) depending on the width of the frequency range (Hz) of the cochlear implant.
As can be seen from this table the highest intelligibility of
words is obtained at a frequency range of 250-6500 Hz. When
the lower boundary of the FR of 250-6500 Hz is shifted to 350 Hz,
the intelligibility of words decreases somewhat. It can be assumed
that when removing the band 250-350 Hz from processing (and
perception) we lose some of the information, but somewhat
increase the intelligibility of speech by increasing the CSS. The
CSS is only slightly increased and the loss of information from the
remote 250-350 Hz band is not compensated by increasing the CSS.
When the upper boundary of the FR 250-6500 is shifted to 8500
Hz, the intelligibility of words decreases significantly. We assume
that this is due to a noticeable decrease in the CSS, which is not
compensated by the information contained in the 6500-8500 Hz
band. Speech intelligibility in FR 70-8500 and 250-8500 does
not differ. No differences in speech intelligibility in these FR can presumably be explained as follows. By adding a band of 70-250
Hz to a FR of 250-8500 Hz, we gain a little information, but we lose
this acquisition due to some decrease in CSS. It is possible that the
acquisition is small. One thing is absolutely certain: a wide FR is
unsuitable for speech perception, since in this case the first two
channels of the implant work almost in vain and, compared with a
narrow FR (250-6500 Hz), the CSS is reduced due to the band of a
channel 12, which is partially needed only for a few HF phonemes.
The obtained results of the study give grounds for a more thorough
investigation in the frequency range of 200-7000 Hz. The search for
a golden mean between the loss of speech information due to the
narrowing of the frequency range of CI and its acquisition due to an
increase in the CSS should be continued.
The above reasoning is also applicable when stimulating a
channels in one run from the first to the last electrode with the
n-of-m strategy for a wide range - up to 10 kHz, since due to the
expansion of the frequency range to 10 kHz, single-channel bands
become wider and thereby the CSS decreases. In addition, there
is an assumption that the wider the channel band, the less the
SPL changes in successive measurements of the instantaneous
spectrum. To test this hypothesis, it is necessary to calculate the
changes of the SPLs in bands of different widths (for example,
300-450 and 300-550 Hz) in time. After the above reasoning,
the issue of disconnecting the implant electrodes should be
considered, which may need to be done for various reasons, for
example, incomplete insertion (how to deal with problems with
the insertion of a chain of electrodes we described earlier [5]) or
mechanical failure of the electrode. If the electrode fails or it is not
inserted, then the frequency band processed in the corresponding
channel falls out of the implant processing and, consequently, out
of the patient’s perception. Such an electrode must be switched
off. After it is turned off, the entire FR is redistributed between the
other electrodes. In this case, the channel selectivity of stimulation
decreases (all bands become wider) and therefore it is necessary to
reduce the previously installed FR, depending on which electrode
has failed. N.B. Disconnecting of electrodes no. 1, 2 or 3 slightly
changes the CSS.
As for the recommendations on disconnecting short-circuited
electrodes, I think that this issue should be considered more closely.
Why? Imagine that 3 and 9 electrodes are short circuited. After
processing the instantaneous spectrum, pulses of the corresponding
amplitude will be created in these channels. When a series of
stimuli is applied from the first to the last electrode to transmit
this instantaneous spectrum, the pulse from the third electrode
will be applied to the 9-th electrode, and the pulse applied to the
9-th electrode will be applied to the 3-rd electrode. Naturally, the
sensations in channels 3 and 9 will differ (tonotopy). With a series
of stimuli to transmit the next instantaneous spectrum, channels 3
and 9 and channels 9 and 3 will also be stimulated simultaneously
and the sensations from each of them will also differ. But what is
important, the sensations from the second series of stimuli of the
instantaneous spectrum will differ from the sensations during the
first series of transmission of the spectrum to these electrodes.
I.e., the information will change both through channels and series,
and it will display changes in the instantaneous speech spectrum.
It should be remembered that the effect of disconnecting a shortcircuited
electrode is the same as when disconnecting a failed
one – the FR is redistributed between the other electrodes and a
decrease in CSS occurs. But in case of a short circuit, the electrode
that transmits the information is switched off. Question: Is it
necessary to disconnect the short-circuited electrode that transmits
information? The above reasoning concerns two short-circuited
electrodes and is not an absolute indication not to disconnect one
of them. To answer the question about disabling one of the two
short-circuited ones, it is necessary to compare the two programs.
Our new method of comparing programs, successfully applied by us
on a model of implants with different FR, can be used to compare
programs in CI patients with one of the two short-circuited
electrodes disconnected and not disconnected. The question of
three short-circuited and other combinations of electrodes requires
further consideration.
a) To achieve the maximum possible intelligibility of speech, it
is necessary to set the optimal frequency range in the speech
processor. The best frequency range of CI for speech perception
is the golden mean between the loss of low and high frequency
information and value of the channel selectivity of stimulation.
b) The issue of disconnecting short-circuited electrodes requires
closer investigation.
c) A new method of comparing programs, successfully applied on
the CI model, can be used in implanted patients.
The speech signal spectra of many European languages differ
slightly and therefore the optimal FR may be the same in different
countries.
Bio chemistry
University of Texas Medical Branch, USADepartment of Criminal Justice
Liberty University, USADepartment of Psychiatry
University of Kentucky, USADepartment of Medicine
Gally International Biomedical Research & Consulting LLC, USADepartment of Urbanisation and Agricultural
Montreal university, USAOral & Maxillofacial Pathology
New York University, USAGastroenterology and Hepatology
University of Alabama, UKDepartment of Medicine
Universities of Bradford, UKOncology
Circulogene Theranostics, EnglandRadiation Chemistry
National University of Mexico, USAAnalytical Chemistry
Wentworth Institute of Technology, USAMinimally Invasive Surgery
Mercer University school of Medicine, USAPediatric Dentistry
University of Athens , GreeceThe annual scholar awards from Lupine Publishers honor a selected number Read More...
The annual scholar awards from Lupine Publishers honor a selected number read more...